Title: Standard operating procedure for operation and calibration of U.V. Spectrophotometer U.V.1800.
1.
Purpose:
To lay down the procedure for the operation and
calibration of Spectrophotometer
UV1800 ( Make :
Shimadzu ) used in the raw
materials, intermediates and finished
products analysis
in the Lab.
3. Scope:
This procedure shall be applicable for operation and
calibration of Spectrophotometer UV1800 in Lab at Fermenta Biotech Limited.
4. Responsibility: Research
Associate/ Research Scientist .
5.0 Procedure :
5.1 Start
up and operation of instrument:
5.1.1 Before starting the
instrument Ensure, there should be no cuvette or the Sillica Bag in the sample compartment.
5.1.2 After initialization
of Spectrophotometer is completed click O.K. and leave the instrument
for 15-20 minutes
to warm up.
5.1.3 After warm up time, select the
required mode i.e. photometric or spectrum by clicking the icon.
5.1.4 Fill the Wavelength in the
photometric mode and wavelength range in the spectrum mode.
5.1.5 Click Baseline for Baseline
correction in spectrum mode and click Auto zero in Photometric method.
5.1.6 Rinse and fill both the cuvettes
with blank solution wipe the outer surface with the tissue paper and place it in respective position in sample
compartment.
5.1.7 Now
again click Baseline for Baseline correction in spectrum mode and click Auto
Zero for making the reading zero of
Blank in photometric mode.
5.1.8 Take
out the front cuvette rinse and fill the cuvette with Test sample, place the
cuvette in the sample compartment.
5.1.9 Click the start, in the spectrum
mode spectrum will be displayed and in the photometric mode. Absorbance will be displayed on the screen.
5.2 Switch off the instrument:
5.2.1 After Analysis take out cuvettes from sample
compartment clean them and keep in cuvette box.
5.2.2 Click Return then click mode, a message
will appear on screen 'current data not saved do you continue, if you want to
exit the application, click O.K, UV probe file will close, and will return to
mode menu.
5.2.3
Switch off the UV Spectrophotometer.
5.3 Calibration of the UV-Spectrophotometer
5.3.1 Instruments
and apparatus:
Analytical weighing balance
Volumetric flask
Pipette
5.3.2 Reagents:
Potassium
dichromate (K2Cr207).
Sulphuric acid A.R
Grade.
Potassium chloride.
Toluene.
Hexane
RO water
5.3.3 Calibration Procedure: Transfer
accurately weighed about 60 mg K2Cr207 previously
dried at 120ºC in 1.0 liter volumetric flask. Add distilled water to dissolve,
add 0.27 ml of Sulphuric acid and make up to 1.0 liter with RO water.
5.3.4 Control of Absorbance: Switch on
the instrument.
5.3.4.1 Initialization
starts.
5.3.4.2 After completion of initialization choose spectrum mode to
check spectrum.
5.3.4.3 Go to method and select the wavelength range 400 nm to 230nm
then enter.
5.3.3.4 Start base line
correction.
5.3.3.5 Fill the both
cuvettes with the blank solution. Place the cuvettes in their respective
position in the instruments.
5.3.3.6 Zero the reading by pressing start baseline
correction.
5.3.3.7 Take out front cuvette and fill with
the 0.006 % K2Cr207 after rinsing put into
respective positions and click start.
5.3.3.8 Then spectrum will
come on the screen then click operation and point peak.
Fill wavelength
350nm, 313nm, 257nm and 235nm.
5.3.3.9 Then absorbance
of above said
wavelength will come
on screen and
take out the print and calculate as under.
Calculation: Absorbance value x1000
A (1%, 1 cm) = -----------------------------------------
Weight of K2Cr207 in gm x100
Wavelength A (%, 1cm) Maximum tolerance
235 124.5 122.9
to
126.2
257 144.0 142.4
to
145.7
313 48.6 47.0 to 50.3
350 106.6 104.9 to 108.2
5.3.5 Limit of stray light:
5.3.5.1 The absorbance of
a 1.2% w/v solution of potassium chloride with a path length of 1.0 cm should
be more than 2.0 at 200nm. When compared with water as reference liquid.
5.3.6
Resolution Power:
5..3.6.1Prepare 0.02% v/v solution of Toluene in
Hexane and take the spectrum from 220nm to 300 nm against water as a blank. Note the absorbance at 269 nm and 266nm
and calculate the resolution power
as under.
Absorbance at 269nm
-------------------------------
= Theoretical ratio should be not less than 1.5
Absorbance at 266nm
5.3.6.2
After calibration is over, record the results.
6.0 Abbreviations:
|
gm |
: |
Grams |
|
ml |
: |
milli liter |
|
nm |
: |
nano meter |
|
mg |
: |
milli gram |
|
UV |
: |
Ultra Violet |
|
cm |
: |
Centimeter |
|
°C |
: |
degree Celsius |
|
v/v |
: |
volume/volume |
7 .Records:
|
Sr.No. |
Name of the Document |
Rev No. |
|
1 |
UV Spectrophotometer Report |
01 |
8. References:
: ICH Guidelines Q7 and
spectrophotometer manual.
No comments:
Post a Comment