Sunday, 1 October 2023

Guidelines for Validation of Analytical Method for Cleaning Validation.

 

1.       OBJECTIVE

To lay down the procedure for validation of Analytical method for cleaning validation in equipment surface samples for cleaning assessment.



2.                   SCOPE

This procedure is applicable for validation of Analytical method for cleaning validation by swab method for the product chosen as worst case and manufactured at different facility of Ajanta Pharma Limited.



3.                   RESPONSIBILITY

3.1               Assistant Manager or Designee Analytical Development Laboratory (ADL):

    Preparation and execution of SOP.

3.2               Head of Department (HOD) or designee ADL:

     To check the SOP and to ensure activities are performed as per SOP. 

3.3               Head or Designee of Research Centre (RC):

     To review and authorization of SOP.

 

4.                   PROCEDURE

Disregard procedure for evaluation in chromatographic analysis. Maximum allowable carryover (MACO) limit shall be provided by plant.



1.       On the basis of limit of swab (MACO limit) the critical limit level concentration (CLLC) shall be calculated as example given below and validation shall be carried out using a specific written and authorized validation protocol considering critical limit level concentration as 100%.

Examples to transform the acceptance criteria value for swab in concentration units (µg/ml) is          presented as

1.       Equipment surface to be swabbed is 5cm X 5cm = 25cm2

2.       Amount of drug from the swabbed are (value provided from location side) = XYZ µg/swab.

3.       Sampling volume e.g. 5mL, 10mL, 20 mL…..100mL

4.       Drug concentration at acceptance level = 250 µg/25 mL = 10 µg/mL. Hence 10 µg/mL is the critical limit level concentration (CLLC).

3.                   Materials required:

1.       Swab TX 714A Large alpha swab

2.       Sampling Template 1.5 mm thick white PVC sheet with 5cm x 5cm area cut on its surface, 5cm x 5cm coupon.

3.       SS Plate 316 or any other material as per requirement such as sieves, roller etc.

4.                   Swabbing procedure

1.       Wet the swab with solvent given in respective protocol. Hold the swab at once corner and wipe the surface of the coupon eight times using sampling template of size 5 x 5 cm in paint brush action as shown in figure 1 and 2, so that whatever amount recovered remains in the swab only. Dip the swab in the glass beaker or conical flask containing or measure volume of 25 ml of diluent or water and remaining sample treatment as per the procedure given in the protocol.

                  

2.                   Calculation:

1.       For Swab Sample:

2.       Amount of drug in µg/cm

Area of drug in spl. Preparation           Wt. of std.              std.dil.1           std.dil.2

--------------------------------------- X   ------------------ X ----------- X ------------- X              Area of drug in std. Preparation                dil.                       dil.1                  dil.2 

 

Sample volume i.e. 25, 50 or 100 mL           1

--------------------------------------------- X   --------- =                      mg/swab

                        25*                                    RCF

Where 25* = represents the sample surface area (cm2)

RCF = represents the recovery correction factor for the swabbed surface obtained from   method validation

5.                   Experimental Design:

1.       Following parameters has to be evaluated for cleaning validation by HPLC method and following general criteria should be consider, if not mentioned in individual protocol.

1.       System suitability

2.       Specificity

3.       Limit of Detection

4.       Limit of Quantitation

5.       Linearity

6.       Range

7.       Recovery

8.       Precision

9.       Intermediate Precision (Ruggedness).

2.       System suitability:

1.       Experimental Design

       Inject five/six replicate injections of Standard preparation

       Acceptance criteria

1) % RSD for peak area response of replicate standards : NMT 2.0% for five  replicate injections and NMT 5.0% for six replicate injections     

       2) Theoretical plates: NLT 2000

       3) Asymmetry: NMT 2.0

3.                   Specificity:

1.       Experimental Design

      Diluent, Placebo and swab interference

      Acceptance criteria

1) There should not be any interference of peak from blank, placebo and swab at the retention time of main peak.

2) Peak purity for the main peak in Standard preparation and Spiked from coupon and swab should not be less than 0.99.

4.                   Limit of Detection:

1.       The Limit of Detection can be determined by any of the method given below.    

Method I

Experimental Design

Perform the linearity and derive the curve by graph (Analyte concentration VS Response) and calculate,

         3.3  ×   σ

LOD   = -----------------

         S

Where, σ = the residual standard deviation of the regression line from calibration curve

S = the slope of the calibration curve.

Acceptance Criteria

Report the value of detection limit.

                                        Method II

   Experimental Design

The Limit for Detection is established by signal-to-noise ratio (S/N) obtained from baseline noise using following formula.

2H

S/N =      ---------

h

                                                                        Where,

                                                                        S/N = Signal to noise ratio

                                                                        H = Height of the peak of interest in mm.

h = Height of the noise in mm.

Acceptance criteria:

1) Report the value of detection limit.

2) Limit of Detection (S/N Ratio) should be about 2 to 3.

5.                   Limit of Quantitation

1.       The Limit of Quantitation can be determined by any of the method given below.

Method I

Experimental Design

Perform the linearity and derive the curve by graph (Analyte concentration VS Response) and calculate,

10 ×   σ

LOQ   =   -------------

S

Where, σ = the residual standard deviation of the regression line from calibration curve

S = the slope of the calibration curve.

Acceptance Criteria

1.       Report the value of Quantitation limit.

2.       RSD of area should not be more than 10.0%.

                                        Method II

   Experimental Design

The Limit for Quantitation is established by signal-to-noise ratio (S/N) obtained from baseline noise using following formula.

2H

S/N =      ---------

h

                                                                        Where,

                                                                        S/N = Signal to noise ratio

                                                                        H = Height of the peak of interest in mm.

h = Height of the noise in mm.

Acceptance criteria:

1) Report the value of Quantitation limit.

2) Limit of Quantitation (S/N Ratio) should be about 10.

3) RSD of area should not be more than 10.0%.

6.                   Linearity and range:

1.       Experimental Design

1.       For the establishment of linearity and range, a minimum 5 concentrations of standard in the specified range as per method validation protocol. Linearity is evaluated by visual inspection of a plot of signals as a function of analyte concentration. Inject first level and last level six times and remaining all the other levels in triplicate. Linearity shall be at LOQ level to 150% of critical limit level concentration.

2.       Acceptance criteria:

The correlation coefficient should not be less than 0.99. The relative standard deviation at each level should not be more than 10.0% for LOQ level and 5.0 % for all the other levels. Report the value of slope of regression line, y-intercept and residual sum of square.

7.                   Accuracy (Recovery):

1.       Experimental Design:

1.       A known amount of working standard solution containing placebo is spike on the swab or on the coupon surface (5cm X 5cm) or on any other material as per requirement ( 5cm X 5cm) such as sieves, roller etc. and evaporate it as per the procedure given in the protocol. Swab the surface and calculate the recovery. Accuracy shall be at LOQ level to 150% of critical limit level concentration.

2.       Acceptance criteria:

Recovery at each level, mean recovery and overall mean recovery from swab should be between 90.0% and 110.0% with %RSD should not be more than 5.0%.

Recovery at each level, mean recovery and overall mean recovery from coupon surface should be between 80.0% and 120.0% with %RSD should not be more than 10.0%.

Recovery at each level, mean recovery and overall mean recovery from any other surface except swab and coupon should be between 70.0% and 130.0% with %RSD should not be more than 15.0%.

8.                   System precision:

1.       Experimental Design:

1.       Inject five replicate injections of Standard preparation

2.       Acceptance criteria

% RSD for peak area response of replicate standards: NMT 5.0%.

Theoretical plates: NLT 2000

Asymmetry: NMT 2.0

9.                   Method precision:

1.       Experimental Design:

1.       Carry out minimum six determinations of Recovery from coupon surface at Critical limit level concentration.

2.       Acceptance criteria

%RSD of recovery of six replicate determinations should not be more than 10.0%.

The recovery should be between 80.0% to 120.0%.

10.               Intermediate precision:

1.       Experimental Design:

1.       Carry out minimum six determinations of Recovery from coupon surface at limit level concentration on different days, by different analyst and using different instruments.

2.       Acceptance criteria

%RSD of recovery of six replicate determinations should not be more than 10.0%.

%RSD of recovery of twelve replicate determination (Method precision and Intermediate precision) should not be more than 10.0%.

The recovery should be between 80.0% to 120.0%.

5.                   RELATED DOCUMENTS

Nil



6.                   REFERENCES

Nil

No comments:

Post a Comment

Operation, Calibration, Cleaning and Maintenance of UV Cabinet

  1.0 OBJECTIVE  To lay down the Procedure for Operation, Calibration, Cleaning and Maintenance of UV ...